The CompAir Elite is on, but nebulizes weakly. What would cause this?
Make sure the nebulizer kit is seated correctly in the Main Unit of the NE-C21 and that the air tubing connection is free of obstruction. The nebulizer should be in an upright position during use. Do not tilt beyond 30ยบ. Check whether the nebulizer is empty. You may need to clean or replace the nebulizer kit. Check the air filter to see if it’s dirty or obstructed.
Make sure the nebulizer kit is seated correctly in the Main Unit of the NE-C21 and that the air tubing connection is free of obstruction. The nebulizer should be in an upright position during use. Do not tilt beyond 30. Check whether the nebulizer is empty. You may need to clean or replace the nebulizer kit. Check the air filter to see if it’s dirty or obstructed.