Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

On the VR2240 why am I seeing the message “no con in5” that locks theinstrument?

locks message seeing
0
Posted

On the VR2240 why am I seeing the message “no con in5” that locks theinstrument?

0

The condition occurs if the user presses the In/Idn button inadvertentlywhile the instrument is in Loop Zs/Ik mode on the rotary switch. Themessage “no con in5” comes up on the display. The instrument will nolonger measure Loop Impedance on the Zs/Ik position. This condition canoccur on firmware releases 1.34 to 1.38 and may occur on subsequentreleases, until it has been verified that the condition has been eliminated.