Is the use of a surrogate urethra an option in prostate high-dose-rate brachytherapy?
PURPOSE: To investigate the accuracy and the dosimetric consequences of substituting a surrogate urethra assumed to be at the geometric center of the prostate, in place of the true urethra when using high-dose-rate (HDR) brachytherapy for the treatment of prostate cancer. METHODS AND MATERIALS: One hundred prostate cancer patients treated with HDR brachytherapy constituted the study group. A pre-plan was made with the urethra visualized. The true urethra was defined, and a surrogate urethra was placed at the geometric center of the prostate. The distance between the two urethras was measured. The deviation was evaluated at the base, middle, and apex. To evaluate the dosimetric consequences for the true urethra when using a surrogate urethra, two different dose plans were made: one based on the true urethra and one based on the surrogate urethra. The dose-volume histograms for the true urethra were analyzed. RESULTS: The deviation between the true urethra and the surrogate urethra was g