Is the Tensys® T-Line® System accurate relative to an A-Line?
A. The T-Line® has been validated against a contralaterally placed invasive A-Line reference and has been found to meet the standards as specified in the AAMI SP-10:2002 Accuracy Specification 4.4.5.2.B. More specifically, this validation testing confirmed a mean difference between T-Line® and A-Line values not exceeding 5 mm Hg with one standard deviation not exceeding 8 mm Hg for systolic, mean and diastolic values.