Is the GraftConnector a valid alternative to running suture in end-to-side coronary arteries anastomoses?
GROUND: An animal study was carried out to compare long-term patency rates of coronary anastomoses performed with the GraftConnector versus running suture technique. METHODS: 10 sheep, 45 to 55 kg, underwent off-pump coronary artery bypass grafting (right internal mammary artery to left anterior descending artery). In 5 animals, the anastomosis was performed with a GraftConnector and in 5 animals with 7-0 running suture. Intraoperative fluoroscopy and a fluoroscopic control at 6 months were performed. After 6 months, the animals were sacrificed and the anastomoses were examined histologically. RESULTS: All animals survived at 6 months with 100% anastomosis patency rates in both groups. In the GraftConnector group, the anastomosis diameter at 6 months fluoroscopy was 118% of native left anterior descending artery versus 97% of the control group. Luminal anastomotic width at histology was 1.7 +/- 0.2 mm in the device group versus 1.6 +/- 0.1 mm in the control group. Mean intimal hyperpla