Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Is the Agilent 16910A state/timing card supported?

0
Posted

Is the Agilent 16910A state/timing card supported?

0

A. Yes and No. The 16910A will work with the FuturePlus HyperTransport analysis probes when Turbo Mode is not required or in Turbo Mode but only as a single machine (8 bit or 16 bit Transmit OR Receive). In order for simultaneous capture of TX and RX two separate state machines are required. Let’s clear up any confusion. When the 16910A modules are placed in turbo mode a pod pair is used for each machine to enable this function. In order for two machines to be in Turbo Mode two pod pairs would be used leaving only the 1 remaining pod pair, which is not enough for 2 machines. Each machine requires a minimum of 1 pod pair. If you have or are considering using an Agilent 16910A state/timing module with a FuturePlus HyperTransport analysis probe, please contact us and let us verify the correct configuration for your application.

Related Questions

What is your question?