Is it possible to rapidly switch its sample input between two streams, to allow control of both independently?
This multiplexing is not recommended. Normally one SCM is required for each independent coagulated water stream. The instrument’s response to significant changes in SC can take more than 10 minutes to settle. Our Accufloc SCM has advanced signal processing which makes it one of the fastest responding SCMs available, however as the sensor becomes more worn the response gets slower. Use with multiple switched streams will result in significant interference between the two readings, unless a very long time is allowed in between, which would make it useless for online control.