Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Is it possible to operate the Centrifuge 5702 in the cooling room?

0
Posted

Is it possible to operate the Centrifuge 5702 in the cooling room?

0

Yes, the Centrifuge 5702 may be operated in the cooling room. When moving the centrifuge from the cooling room to a normal lab room, the centrifuge must either run warm for half an hour in the cooling room or warm up for at least 3 hours in the lab room without being connected to the power supply. This is to avoid damage caused by condensation.