Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

If I obtain a good fit between an experimental voltammogram and a simulation, does this mean that the selected mechanism and parameters are correct?

0
Posted

If I obtain a good fit between an experimental voltammogram and a simulation, does this mean that the selected mechanism and parameters are correct?

0

No. A complicated experimental voltammogram can probably be reasonably well described by a number of different mechanisms. It is important to examine the changes in the voltammograms over a range of scan rates and concentrations. Even if a series of experimental voltammograms run over a range of scan rates are well fit by a single set of simulation parameter values, the sensivity of the fit to each of the parameters of interest must be explored.