Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

I want to add an auxiliary sensor to my CTD (SBE 9plus, 16, 16plus, 16plus-IM, 16plus V2, 16plus-IM V2, 19, 19plus, 19plus V2, or 25). Assuming the auxiliary sensor is compatible with the instrument, what is the procedure?

Auxiliary ctd SBE sensor
0
Posted

I want to add an auxiliary sensor to my CTD (SBE 9plus, 16, 16plus, 16plus-IM, 16plus V2, 16plus-IM V2, 19, 19plus, 19plus V2, or 25). Assuming the auxiliary sensor is compatible with the instrument, what is the procedure?

0

Adding the sensor(s) is reasonably straightforward: • Mount the sensor (Note: If the new sensor will be part of a pumped system, the existing plumbing must be modified; consult Sea-Bird for details.). • Attach the new cable. • Using the appropriate terminal program ‑ Enable the channel(s) in the CTD, using the appropriate instrument command. • Using SEASAVE V7, SEASAVE-Win32, or SBE Data Processing ‑ Modify the CTD configuration (.con) file to reflect the new sensor, and type in the calibration coefficients.