I have problems with a variable Research pipette which deviates significantly from the limit values. How can I establish whether this is a measuring error or a problem with the pipette?
“Dismantle the pipette into its individual parts (to do this, see operating instructions) and check whether there is any contamination. If so, you can wipe this off with a cloth and reassemble the pipette. Then recalibrate your pipette and follow the instructions in our SOP as you do so. Values which deviate significantly from the limit value should now no longer occur. If there is no contamination, please send the pipette to our Pipette Service department.