Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

I am using the Humidifier System, however, it seems that humidification does not work properly since evaporation still takes place. What can be the reason?

0
Posted

I am using the Humidifier System, however, it seems that humidification does not work properly since evaporation still takes place. What can be the reason?

0

For an efficient humidification the condensation vessel of the Humidifier System should also be filled with 0.5-1 cm of distilled water. Additionally, the use of a specific FoilCover is recommended for best results. When using the Humidifier System, the ventilation setting at the CTI-Controller 3700 digital or the Heating Unit should be increased. The system reduces the air flow of the circulating air stream, which has to be compensated. If the air flow in the incubation system is too low, the humidified air is not transferred effectively to the incubator.

What is your question?

*Sadly, we had to bring back ads too. Hopefully more targeted.