I am testing an ELT with separate 121.5 and 406 output connectors. How do I get the results printed on one Test Report?
Connect to the 121.5 transmitter first, then tap the Receive button. Once the Beacon Tester gives the indication that the 121.5 signal has been received, switch the cable to the 406 transmitter without stopping the test. Once the 406 burst is received on the Beacon Tester, the measurement will be complete.