I am making high-power measurements. It there a power limitation on electronic calibration or ECal?
The maximum power rating for ECal modules is either +10 or +20 dBm (see Table below). The ECal module also has a minimum power requirement for auto-orientation (not calibration, but orientation). If the power level at the module is less than 18 dBm, the user has to tell the analyzer the orientation of the ECal module. Simply unselect the automatic orientation check box and manually indicate to the analyzer how the ECal module is connected. Auto-orientation means that the network analyzer determines how port 1 and 2 are connected to ports A and B of the ECal module.