Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How will an inspector inspect a CR unit being used under the certificate extension program?

0
Posted

How will an inspector inspect a CR unit being used under the certificate extension program?

0

Using his/her laptop computer, the inspector will download for inspection the record for a mammography unit(s) that uses CR in the same manner as any other FFDM unit, and then open the applicable inspection procedure screens. If a unit(s) is being used for both S-F and CR on the same physical unit(s), the inspector will need to download a separate S-F record, as well. If an inspector determines that CR is being used on an S-F unit(s) for which there is no certificate extension approval, the inspector will: • add a new record for the CR unit(s) into the database (e.g., with the same model, description, and serial number as the S-F unit(s) but with an image receptor type of CR); • mark it as unaccredited (resulting in a Level 1 or 2 inspection observation); • answer the mammography equipment evaluation (MEE) question and all of the applicable questions in the QC records and Survey Report sections accordingly; and, • instruct the facility that it should stop using the unit(s) until it app