Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How Should A pH Electrode Be Installed?

Electrode installed pH
0
Posted

How Should A pH Electrode Be Installed?

0

All pH electrodes (and DO sensors) can be mounted vertically, into a vessel headplate. Electrodes and sensors can also be installed on large vessels, through a side port. For side-entry installations, the probe must be at least 15° above horizontal to insure proper operation. Remember, there is liquid inside all pH electrodes and DO sensors. This slight 15° incline is enough to prevent a bubble from lodging itself at the tip of the sensor, preventing the sensor from operating properly.