How many units of blood can I safely infuse through the Thermal Angel TA-200?
The Thermal Angel TA-200 has been evaluated utilizing standard blood administration sets. The filter on these sets is the limiting factor in determining the maximum number of units allowed through one device. Laboratory tests have demonstrated three (3) to four (4) units of blood can be successfully administered before the filter is occluded. At this point, the administration set should be discarded, as should the Thermal Angel TA-200. Note: Multiple TA-BCE Batteries may be needed for warming of this much blood.