Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How long does a Kaye Teflon probe need to stabilize, going from room temperature into a HTR/LTR at 121°C?

0
10 Posted

How long does a Kaye Teflon probe need to stabilize, going from room temperature into a HTR/LTR at 121°C?

0
10

If your temperature reference is already stable at 121°C, the thermocouple should stabilize within three to five minutes after inserting thermocouples at ambient. A probe with a leak in the insulation (moisture in the tip), will not stabilize in this time frame. If you use Kaye’s software for automated sensor calibration, it automatically monitors probe stability and captures the data until stability is achieved.