How is the accuracy calculated for CR mode?
1) Convert Rset to mS units > mSset (Resistance value in mS units). 2) Calculate %error in mS units. 3) Calculate mSset + %error + Offset. 4) Calculate mSset + %error – Offset. 5) Convert both values back to resistance values to obtain accuracy range. Is there an external method for shutting off the eLoad input? Short the remote inhibits (RI) rear terminal connectors. Also, an available TTL fault signal will trigger into an active high state, in the case of any fault conditions. Does the water-cooled eLoad offer an anti-condensation feature? Yes. The eLoad incorporates an automatic anti-condensation, liquid cooling system. This is accomplished via a water valve controlled by the load’s operating temperature.