How is Mean Kinetic Temperature (MKT) calculated in the TTMD application?
• Using this formula: Kevin(I) = C(I) + 273, convert each Celsius value to Kelvin(I), where I = 1,2, .., .N, and N is the number of points for the monitor. • Calculate PointSum(I) = Exp(-9982.68/Kelvin(I)) for each point. • Calculate PointSum = ∑PointSum(I), where I = 1,2,…, N • Calculate MKTK = -9982.68/Ln(PointSum/N) (This is MKT in the Kelvin temperature scale.) • Convert MKTC using the following formula: MKTC = MKTK – 273 (This is MKT in the Celsius temperature scale. This value displays on the Summary View page.) Q: How do I view humidity data in the multi-record graph view? A: This answer applies to any dual-sensor TempTale monitor. While viewing the multi-graph, select Secondary in the Sensor panel located at the bottom right of the Graph view. Q: How do I make the dates smaller on the TTMD graph printouts? A: Change the font size on the graph by clicking on View from the Menu Bar. Select Graph Options from the drop-down menu. (Note: You must be viewing the Graph to execute Gr