How good is the ACR accreditation phantom for assessing image quality in digital mammography?
RATIONALE AND OBJECTIVES: The purpose of this study was to evaluate the American College of Radiology (ACR) accreditation phantom for assessing image quality in digital mammography. MATERIALS AND METHODS: Digital images were obtained of an ACR accreditation phantom at varying mAs (constant kVp) and varying kVp (constant mAs). The average glandular dose for a breast with 50% glandularity was determined for each technique factor. Images were displayed on a 5 mega-pixel monitor, with the window width and level settings individually optimized for viewing the fibers, specks, and masses in the ACR phantom. Digital images of the ACR phantom were presented in a random manner to eight observers, each of whom indicated the number of objects visible in each image. RESULTS: Intraobserver variability was greater than interobserver variability for the detection of fibers and specks, but the reverse was true for the detection of masses. As the mAs increased, the number of fibers visible increased fro