Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How fast should I deflate my sphyg during measurement?

deflate fast measurement sphyg
0
Posted

How fast should I deflate my sphyg during measurement?

0

The AHA recommends a deflation rate of 2 – 3mmHg per second. During routine measurement, the system is typically inflated to about 180mmHg and deflated to about 60mmHg to a range of 120mmHg (the actual range is determined by the patient’s expected blood pressure). At a deflation rate of 2-3mmHg/sec, it should take 40 to 60 seconds to deflate the cuff correctly. Excessive deflation rates are one of the most common causes of measurement error – often distorting readings by 10mmHg to as much as 20mmHg. When the deflation rate is too quick, the observer’s reaction time – much slower than the rapid deflation rate – may result in gross underestimation of blood pressure. On the other hand, because of a phenomenon known as mercury lag, excessive deflation COULD result in an overestimation of blood pressure when measured on a mercurial sphyg.