How does the Needless device affect the Groshong* valve?
Positive pressure needleless access devices do not affect the function of the Groshong* valve. Negative pressure devices, such as the CCU Medica Product, Coeve*, or the Baxter Clearlink, product may cause the Groshong* valve to remain in the open position which could lead to blood reflux and occlusion. QUESTION: Does clamping the extension tubing that is placed on Groshong* PICC lines affect the valve? Will clamping the extension set cause the valve to remain open? What about if it has a needless system on it? ANSWER: Estension leg clamps will not affect the Groshong* valve, however, clamps are not needed with the Groshong* valve. No (zero) clamps used in connection with a Groshong* PICC and a positive pressure needleless access device will have no affect on valve function as long as the clamp is closed following disconnection of the syringe.
Positive pressure needleless access devices do not affect the function of the Groshong* valve. Negative pressure devices, such as the CCU Medica Product, Coeve*, or the Baxter Clearlink, product may cause the Groshong* valve to remain in the open position which could lead to blood reflux and occlusion.