Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How does the Hydra need to be configured to be used with MET/TEMP II?

configured hydra ii Met Used
0
Posted

How does the Hydra need to be configured to be used with MET/TEMP II?

0

A. When using a Hydra Series II Dataglogger with MET/TEMP II, you must enable and configure each channel manually, before starting a calibration, to read the appropriate scale. When a calibration test starts, MET/TEMP II will send a command to the Hydra to enable scanning. The Hydra MUST be left scanning all enabled channels for the duration of the test for MET/TEMP II to be able to take readings on all enabled channels. When the test is complete, MET/TEMP II turns the scan function off.