How does the Hydra need to be configured to be used with MET/TEMP II?
A. When using a Hydra Series II Dataglogger with MET/TEMP II, you must enable and configure each channel manually, before starting a calibration, to read the appropriate scale. When a calibration test starts, MET/TEMP II will send a command to the Hydra to enable scanning. The Hydra MUST be left scanning all enabled channels for the duration of the test for MET/TEMP II to be able to take readings on all enabled channels. When the test is complete, MET/TEMP II turns the scan function off.