How do you run controls on the Cholestech LDX System?
Detailed information on how to run controls can be found on the Cholestech LDX Assay Sheet shipped with the product. Below is a brief summary of this information. Controls should be run only when the material is at room temperature Make certain that the lot number of the controls corresponds to the correct assay sheet when comparing expected results. Mix each vial by gentle inversion 7-8 times Use the appropriate MiniPetâ„¢ Pipette and tips to dispense the control material onto the test cassette. Use a new tip for each control level Follow the direction for the test cassette procedure being used. Controls are to be tested in the same manner a patient’s sample would be tested.