How do two 9.3 mm drill holes preserve the lateral wall when one 16 mm drill hole from a single axis design causes fracturing and lateral wall collapse?
Bone preservation anterior and posterior with the two smaller diameter drill holes will prevent fracturing of the greater trochanter and subsequent collapse. The one, larger diameter drill hole sacrifices necessary bone anterior and posterior which frequently contributes to fracture collapse and distal fragment medialization.