Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How do I set up the pulser for the best MCA linearity measurement?

0
Posted

How do I set up the pulser for the best MCA linearity measurement?

0

Best results are accomplished by multiple sweeps at the shortest ramp time (90 seconds). Set the number of sweeps to 999, giving a run time of just over 24 hours. For unusually demanding needs, the measurement can be repeated for several days. Since any temperature corrections are made between sweeps (not during the ramp), ramp times of 15 minutes may not give the required accuracy. Short ramps allow good random statistics for all channels as long as many sweeps are made.