How do I install a UviClear transparent glass protector?
To replace a worn or degraded UviClear, remove the screws that attach it to the top of the UV transilluminator. Align the new UviClear with the screw holes and tighten all the screws completely. NOTE: If your FOTO/Prep transilluminator has a removable (magnetic) UViClear, do not remove any screws. Merely lift the worn UviClear and set the new UviClear in its place. Exposure to ultraviolet light causes eventual solarization of the UviClear Transparent Protector, which reduces the intensity of transmitted UV light. Replacement of the UviClear is recommended after approximately 50 hours of use.