How do I check the heater resistance on my Dynagage?
A. You will need a voltmeter to check the resistance between pin D and pin E. The correct value for the heater resistance is on the serial number tag for the sensor. Q. What is the maximum number of gages can be connected to one AVRD? A. The number depends on the size of the gages used. The total current used by the sensors must not exceed 5 Amps per an AVRD channel. Refer to the Solar Calculation Chart to find out how many Amps are needed. Enter the number of sensors of each particular type in the spreadsheet, and then the total amperage is computed in cell 18F.
Related Questions
- I gather that the heater resistance should be around 33 ohms. Do resistance values of 37-48 ohms indicate a problem with the sensor, or is that within the normal range?