Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How do I calibrate my Flinn pH meter?

0
Posted

How do I calibrate my Flinn pH meter?

0

If the pH electrode is dry, soak it in tap water for 5 minutes. Add pH 7 buffer solution to a small beaker or vial up to a depth of 1/2 inch. Immerse the electrode in the sample of pH 7 buffer and allow the reading to stabilize. Use a small screwdriver to adjust the pH 7 calibration screw at the top of the meter until the display reads 7.00. Rinse the electrode with distilled water. Add pH 4 buffer solution to a different small beaker or vial and immerse the electrode in the sample. Allow the reading to stabilize. Use the small screwdriver to adjust the pH 4/10 calibration screw until the display reads 4.00.