How closely should the INRatio correlate to my current lab?
The INR value depends on many factors. For more information please see the Technical Bulletins. However as per the ISO 17593 draft titled Clinical laboratory testing and in vitro diagnostic test systems in vitro monitoring systems for anticoagulant therapy: Below INR 2.0, 90% of results shall be within +/- 0.5 INR, and between 2.0 to 4.0, 90% of results shall be within +/- 30%.