Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How can the performance of ORP (redox) Electrodes be checked?

0
Posted

How can the performance of ORP (redox) Electrodes be checked?

0

The performance of an ORP electrode can be determined either by measurement in specially prepared buffer solutions saturated with quinhydrone or with ready-to-use Reagecon Redox Standard (e.g. Redox Standard 250 mV @ 25°C, Cat No. RS250). When the electrode responds properly, calibration is unnecessary. However, the electrode may become contaminated and give erroneous readings. Hence, it is desirable to check the performance of redox electrodes using a Reagecon Redox Standard.

Related Questions