Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How can the centrifuge be opened in the case of power failure?

0
Posted

How can the centrifuge be opened in the case of power failure?

0

First, pull the plug and wait until the rotor comes to a standstill (can take up to 5 minutes!). The emergency lid release is located on the bottom of the device: a small, white plastic button is located in the base plate in front of the front right device foot. Remove this and pull the cord vertically downwards. The lid will open. Please ensure that the cord retracts completely into the housing when you later close the lid. Press the plastic button back into the base plate.