Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How are ECT sensors calibrated?

calibrated ECT sensors
0
10 Posted

How are ECT sensors calibrated?

0
10

When a mixture of 2 dielectric materials is to be imaged, ECT systems are normally calibrated by measuring two reference sets of inter-electrode capacitances, CL and CH with the sensor filled with the lower and higher permittivity materials in turn. These values are then used to normalise the subequent capacitance measurements as described in the following FAQ.