Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

During zero calibration, the display shows the summation of the zero gas O2 value plus the present background level and previous background level, correct?

0
Posted

During zero calibration, the display shows the summation of the zero gas O2 value plus the present background level and previous background level, correct?

0

Not really. In Zero Calibration Mode, the unit display the value of the background current of the sensor as expressed as O2 units. (i.e. 75ppb of background current, 100 ppb of background current, etc).