Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

During CPAP TITRATION, are both thermal sensors and nasal pressure sensors required for scoring apnea and hypopnea?

0
Posted

During CPAP TITRATION, are both thermal sensors and nasal pressure sensors required for scoring apnea and hypopnea?

0

The scoring manual does not specifically apply to scoring of respiratory events during CPAP treatment. Alternative methods for verifying flow such as flow output from the CPAP device will be necessary during treatment as nasal pressure and thermal sensor measures may no longer be reliable. R.7.