Does monitoring of pre-/post-dialyzer pressure difference improve efficiency in intermittent hemodialysis?
GROUND: Continuous monitoring of pre-/post-dialyzer pressure difference (DeltaP) is widely used in continuous renal replacement therapies to monitor extracorporeal circuit function. The aim of this study was to verify whether DeltaP may help to identify chronic subclinical worsening of dialysis quality due to incomplete dialyzer clotting in intermittent hemodialysis. METHODS: Nine chronic hemodialysis patients were enrolled in the study and dialyzed twice (high-flux polysulfone dialyzer) with DeltaP and urea-clearance monitoring: the first session with a standard anticoagulation and the second without. To verify whether a visible clotting of the dialyzer precedes or follows a significant DeltaP increase, we checked the dialyzers for the presence of red clots after a saline flush performed when a 50% increase in DeltaP was registered. RESULTS: In the second dialysis session after a 50% increase in DeltaP (documented in 7/9 patients), all dialyzers, after saline flush, showed a visible f