does it detect the pericardial reflection rather than the right atrium?
GROUND AND OBJECTIVE: Although electrocardiography (ECG) guidance of central venous catheters (CVCs) is traditionally thought to detect the entrance into the right atrium (RA), there is little evidence in the literature to confirm this. We previously observed a high incidence of left-sided CVCs abutting the wall of the superior vena cava (SVC), even when the catheters were advanced past the point of increased P-wave amplitude. Our hypothesis was that this ECG amplitude signal is actually detecting the pericardial reflection rather than the RA. The goal of the study was to position catheter tips under ECG guidance outside the RA. METHODS: One-hundred central venous triple-lumen catheters inserted either via the right or the left internal jugular veins, respectively, were analysed in cardiac surgical patients. The position of the catheter tip was ascertained by ECG. METHOD A: A Seldinger guide-wire in the distal lumen served as exploring electrode, the respective insertion depth was reco