Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Does an alteration of dialyzer design and geometry affect biocompatibility parameters?

0
Posted

Does an alteration of dialyzer design and geometry affect biocompatibility parameters?

0

The aim of the study was to assess the biocompatibility profile of a newly developed high-flux polysulfone dialyzer type (FX-class dialyzer). The new class of dialyzers incorporates a number of novel design features (including a new membrane) that have been developed specifically in order to enhance the removal of small- and middle-size molecules. The new FX dialyzer series was compared with the classical routinely used high-flux polysulfone F series of dialyzers. In an open prospective, randomized, crossover clinical study, concentrations of the C5a complement component, and leukocyte count in blood and various thrombogenicity parameters were evaluated before, and at 15 and 60 min of hemodialysis at both dialyzer inlet and outlet in 9 long-term hemodialysis patients using the FX60S dialyzers and, after crossover, the classical F60S, while in another 9 patients, the evaluation was made with the dialyzers used in reverse order. The comparison of dialyzers based on evaluation of the grou

Related Questions

What is your question?

*Sadly, we had to bring back ads too. Hopefully more targeted.

Experts123