Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Do the CA-CALC™ Series CA-6200 Combustion Analyzers calculate CO undiluted (air free)?

0
Posted

Do the CA-CALC™ Series CA-6200 Combustion Analyzers calculate CO undiluted (air free)?

0

When the O2 reference is turned on in the menu setup and is corrected to reference 0% Oxygen, the new recalculated values are said to be undiluted or air free values. To set up the instrument to display undiluted values, go into the MENU In the instrument setup screen, turn the O2 reference ON and set the O2 reference to 0%. Regulatory authorities often require gases to be corrected to other reference oxygen percentages, often 15% or 3%.