Do TAI sensors detect stoppage of oxygen flow?
Our sensors measure the partial pressure of oxygen and are, for most purposes, flow insensitive. What is important is the partial pressure of oxygen in the no flow condition, which is a function of leakage paths, oxygen level, and volume of gas in contact with the cell, and the oxygen profile of the gas that leaks into the measurement system with no flow, and the rate of this leakage. These factors can vary significantly from system to system.