Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Do pH buffers and filling solutions have a shelf-life?

0
Posted

Do pH buffers and filling solutions have a shelf-life?

0

The typical shelf-life for pH buffers and filling solutions is 2 years unopened and 6 months open. For best results, the pH buffer bottles should be sealed promptly to avoid carbon dioxide absorption. Back to Top • What is a good pH electrode slope range? The acceptable slope range is 92% to 102%. Slopes below 92% indicate that the electrode may require cleaning or if cleaning does not help, the electrode should be replaced. Slopes above 102% indicate that the pH buffers are contaminated. General Note: Most Orion pH meters display the slope as a percentage of the theoretical slope (59.12 mV/pH unit). For example, a 98.5% slope is equivalent to a slope of 58.27 mV/pH unit. Back to Top • Do I need an Automatic Temperature Compensation (ATC) probe? The most common cause of error in pH measurements is temperature. The slope of a pH electrode is highly dependent on temperature, and pH buffer values and sample values change with temperature. For the most accurate results an ATC probe is alwa