Concerning the requirement to include information about changes on labels to required information in material transfer reports, what is meant by “the required information”?
The Commission is interested in changes to the information specifically required to be included on the label by 10 CFR 32.51(a)(3) and (4) that is also required in the material transfer reports (i.e., information necessary for the tracking of devices). Changes to the device model number, device serial number, isotope, and activity during device repair or maintenance require a corresponding change to the device labeling and require reporting for the quarter in which the labeling change occurred. For activity, this means a new initial activity, not accounting for decay. (NRC Form 653B has spaces for changes to device serial numbers, isotopes, activity, and units.) The report must identify the general licensee, the device, and the applicable changes to the information on the device label. Changes to source serial numbers do not need to be reported unless the same number is also used as the device serial number. [Note: The manufacturer/initial transferor name is that used at the time of th