Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Can we use nitrogen (in other words, 0% O2) to calibrate our Oxigraf X2004 sensor at the low calibration point?

0
Posted

Can we use nitrogen (in other words, 0% O2) to calibrate our Oxigraf X2004 sensor at the low calibration point?

0

You must calibrate the X2000 series with a gas mixture that contains at least 2% O2. Some absorption is necessary in order for the sensor line lock servo to maintain lock. The X2008 in high resolution, low range mode (0-10%) can utilize pure N2 for its low calibration reference gas.