Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Can we measure low level photosynthesis radiation (0,1 μmol/m²s) with the PAR Lite2 and the AMPBOX?

0
Posted

Can we measure low level photosynthesis radiation (0,1 μmol/m²s) with the PAR Lite2 and the AMPBOX?

0

No, 0,1 μmol/m²s input will result in a very small change on the output of the AMPBOX (around 4,00033 mA). This is outside the accuracy of the AMPBOX and the accuracy of the PAR Lite2.