Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Can the E terminal be used to power other non-PACE sensors?

0
Posted

Can the E terminal be used to power other non-PACE sensors?

0

In some cases, yes. The E terminal provides a precise 5 vdc at a maximum current of about 20 milliamps. If you are logging at a fast sample rate (1Hz – 200Hz) the E terminal output is a constant 5 vdc. However, if you are logging at a standard rate (2 secs or slower), the E terminal output is pulsed for a short time just prior to the XR440 sampling its inputs (either 0.025 secs or 0.1 secs, depending on the selected Setup). This is done to insure a long battery life. So even if your non-PACE sensor’s power requirement is less than 20ma at 5vdc, it also needs a very fast warm-up time in order to work with the XR440.