Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Can the Centrifuge 5424 be opened in case of a power failure?

0
Posted

Can the Centrifuge 5424 be opened in case of a power failure?

0

Yes, in case of a power failure, the device can be opened with the help of the rotor key. Before opening the centrifuge, pull the power plug and wait until the rotor stops spinning. The emergency lid release is on the right-hand side of the centrifuge. Remove the plastic cap and insert the rotor key into the nut behind the cap. Turn the rotor key counter-clockwise. The centrifuge lid will be unlocked and can be opened. Then remove the rotor key and replace the plastic cap.