Can LogTag recorders be calibrated?
There is a common misconception about the meaning of certain terminology used in connection with calibration. Calibration is the process of checking a given instrument against a reference with a calibration table of values and errors being generated. Any instrument can be calibrated. What clients usually want is the ability to “adjust” a LogTag. Adjustment (or recalibration as it is sometimes called) is when an instrument is adjusted to read in accordance with a reference instrument. The readout of the instrument after the adjustment is the same as the reference instrument All LogTags can be re-calibrated (adjusted) with an application called “LogTag Calibrate”. Please discuss this requirement with your local distributor.The LogTag is designed to operate to the published accuracy over the typical 2 year life provided it is operated in the rated environment defined in the product specification. Temperature calibration therefore should not be required, unless it is a third party requirem