Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Can I use narrow range equations like 49, 51, 53 with the SRP5 or SRP6 logger?

equations Logger narrow range
0
Posted

Can I use narrow range equations like 49, 51, 53 with the SRP5 or SRP6 logger?

0

No. The correct equation numbers for the SRP5 and SRP6 loggers are: • J type thermocouple: Equation 50 • K type thermocouple: Equation 52 • T type thermocouple: Equation 54 • S type thermocouple: Equation 74 Do not use equation numbers 49, 51, 53 and 73 with SRP loggers.