Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Can an air detector replace the Bowie and Dick Test?

air bowie Detector replace
0
Posted

Can an air detector replace the Bowie and Dick Test?

0

No, the air detector cannot recognize entrapped air inside a pack but only monitors condition in the free space of the chamber. The quantity levels at which non condensable (air) gases become critical depend on the goods to be sterilized. For example, a Bowie Dick 7kg linen pack signals an error by means of an inserted indicator sheet or by growth of inserted biological Indicators, if 150-200ml of air has a total inner volume of only 3.14ml. Therefore 0.314ml of non-condensable gases will block 10cm of the tube length to steam penetration, thus preventing effective sterilization of this instrument or tubing at this location.